OAK

Control of voice coil motor nanoscanners for an atomic force microscopy system using a loop shaping technique

Metadata Downloads
Abstract
The voice coil motor nanoscanner has the advantages of large working range, easy control, and low cost compared to the conventional lead zirconate titanate driven nanoscanner. However, it has a small damping problem which causes mechanical vibration. The mechanical vibration reduces the accuracy as well as servobandwidth, which deteoriates the atomic force microscopy (AFM) image of the samples. In order to solve the vibration problem, the loop shaping technique [for vertical (z)] and input prefilter [for lateral (xy)] are applied. Experimental results of the proposed techniques are presented for vertical (z) and lateral (xy) scanner. Finally, the AFM images are provided to investigate its effect. (C) 2008 American Institute of Physics.
Author(s)
Youm, WoosubJung, JongkyuLee, SungQPark, Kyihwan
Issued Date
2008-01
Type
Article
DOI
10.1063/1.2829990
URI
https://scholar.gist.ac.kr/handle/local/17489
Publisher
AMER INST PHYSICS
Citation
REVIEW OF SCIENTIFIC INSTRUMENTS, v.79, no.1
ISSN
0034-6748
Appears in Collections:
Department of Mechanical and Robotics Engineering > 1. Journal Articles