OAK

Improvement of the accuracy of continuous hematocrit measurement under various blood flow conditions

Metadata Downloads
Abstract
We propose an accurate method for continuous hematocrit (HCT) measurement of flowing blood under varying plasma conditions of electrical conductivity, osmolality, and flow rate. Two parameters, namely the hematocrit estimation parameter (HEP) and normalized difference, are proposed to reduce the HCT measurement error. HEP was demonstrated in a previous work. The results of multiple linear regression analysis showed that the two parameters were strongly correlated with the reference HCT measured by microcentrifugation. The measurement error was less than 9% despite significant simultaneous variations in the plasma properties and shear rate. (C) 2014 AIP Publishing LLC.
Author(s)
Kim, MyounggonYang, Sung
Issued Date
2014-04
Type
Article
DOI
10.1063/1.4871706
URI
https://scholar.gist.ac.kr/handle/local/15185
Publisher
AMER INST PHYSICS
Citation
APPLIED PHYSICS LETTERS, v.104, no.15
ISSN
0003-6951
Appears in Collections:
Department of Mechanical and Robotics Engineering > 1. Journal Articles