Ad hoc calibration of interferometric system for measuring nanometer-scale displacements induced by laser ultrasound

Metadata Downloads
Author(s)
Kim, YounggueYoon, TaeilLee, Byeongha
Type
Article
Citation
Optics and Laser Technology, v.181
Issued Date
2025-02
Abstract
Laser Ultrasound (LUS) is commonly used in many fields including thickness measurement and defect inspection. In a conventional LUS system, a piezo-based transducer (PZT) is generally used for detecting the ultrasound echo waves, which requires direct contact with a specimen and thus prolongs measurement time when any lateral scanning is necessary. We present a novel non-contact interferometric system based on a 3 × 3 optical fiber coupler. Even though the 3 × 3 interferometric system works stably at any operating point and allows quantitative measurements, it is generally known that careful calibration is necessary before main measurements. Experimentally, it was observed that the surface displacement, induced by the ultrasound wave of LUS, of a cornea phantom was so minute that averaging was necessary. In this study, we discovered that by using the multiple data sets acquired for averaging, we could obtain the system ad hoc characteristic ellipse without performing the conventional calibration process. Furthermore, by utilizing coherent average we could extract the displacement with a 0.14 nm sensitivity. We could also measure the thickness variation, induced by ocular pressure, of the cornea phantom with a resolution of 4.3 μm by measuring the time of a round trip of the ultrasound wave. This straightforward system, composed solely of a 3 × 3 coupler, is expected to promise a compact and efficient solution to diverse applications. © 2024 The Authors
Publisher
Elsevier Ltd
ISSN
0030-3992
DOI
10.1016/j.optlastec.2024.111779
URI
https://scholar.gist.ac.kr/handle/local/9076
공개 및 라이선스
  • 공개 구분공개
파일 목록

Items in Repository are protected by copyright, with all rights reserved, unless otherwise indicated.