Design and Simulation of a Multi-Receiver Coil Array–Based Magnetic Particle Imaging System without Selection Field Generation

Metadata Downloads
Author(s)
Tahir, Muhammad Umar; Cao, Thanh-Luu; Oh, Seungjun; Yoon, Jungwon
Type
Article
Citation
International Journal on Magnetic Particle Imaging, v.12, no.1
Issued Date
2026-03
Abstract
Conventional magnetic particle imaging (MPI) relies on high-gradient selection fields to create field-free point (FFP) or field-free line (FFL) for spatial encoding; however, these fields require high power and limit scalability. We present a field free region-less (FFR-less), fully stationary MPI configuration that achieves 3-D spatial encoding without mechanical motion. The system integrates a single excitation field with a multi-receiver coil array, providing coverage of a 100 × 100 × 50 mm3 (extendable) field of view. Simulation results confirm nanoparticle localization without selection fields or motion, demonstrating a compact, power-efficient, and scalable path toward human-scale MPI. © 2026 Infinite Science Publishing and Tahir et al.
Publisher
Infinite Science Publishing
ISSN
2365-9033
DOI
10.18416/ijmpi.2026.2603050