OAK

결정론적 측면 장치의 임계 입자 직경에 대한 채널 높이의 영향

Metadata Downloads
Author(s)
Bae JaehyunZhbanov AlexanderYang, Sung
Type
Conference Paper
Citation
2022 마이크로나노시스템학회 추계학술대회
Issued Date
2022-11-17
Abstract
The separation of biological cells or microorganisms in a liquid based on their size by deterministic lateral displacement is widely used in laboratories. The analytical equation for the critical diameter is derived under the assumption that flow between two posts is better described by flow in a rectangular tube than between parallel plates. The height position of the particle is an additional parameter that affects the critical diameter. Preliminary experiments were carried out on the separation of particles in deep and shallow microchannels. This study shows that the critical diameter is not a constant value for a given design but is different on each plane parallel to the top and bottom of the channel. The theoretical model was used to analyze experimental data on the separation of particles larger than 4.2 µm from particles ranging in size from 2.5 to 7.9 µm
Publisher
(사)마이크로나노시스템학회
Conference Place
KO
URI
https://scholar.gist.ac.kr/handle/local/21768
공개 및 라이선스
  • 공개 구분공개
파일 목록
  • 관련 파일이 존재하지 않습니다.