OAK

광학 영상기를 위한 실시간 영상 신호 처리에 관한 연구

Metadata Downloads
Alternative Title
Video Rate Image Signal Processing for Optical Coherence Tomography
Abstract
Optical coherence tomography(OCT) is high resolution imaging system which can see the cross section of microscopic organs in the living tissue. In this paper, we analyze the relation between the light source and the resolution of modulated signal in Michelson interferometer. We construct 1-D OCT signal processing hardware such as amplifiers, filters, and demodulate electronic signals from the photo detector. In order to get 2-D OCT image, the synchronization among optical delay line, sample stage and A/D converter is dealt with. In experiments, we verify analog and digital signal processing blocks which apply to the stacks of glasses. Finally we aquire high resolution 2-D OCT image with respect to the onion tissue. We expect that this result can be applied to the medical instrument through performance improvement.
Author(s)
이창수나지훈이병하
Issued Date
2004-03
Type
Article
DOI
10.5302/J.ICROS.2004.10.3.239
URI
https://scholar.gist.ac.kr/handle/local/18265
Publisher
제어·로봇·시스템학회
Citation
제어.로봇.시스템학회 논문지, v.10, no.3, pp.239 - 248
ISSN
1976-5622
Appears in Collections:
Department of Electrical Engineering and Computer Science > 1. Journal Articles
공개 및 라이선스
  • 공개 구분공개
파일 목록
  • 관련 파일이 존재하지 않습니다.

Items in Repository are protected by copyright, with all rights reserved, unless otherwise indicated.