Optics of Refractometers for Refractive Power Measurement of the Human Eye
- Abstract
- In the field of ophthalmology, many diagnostic instruments based on optical technology have
been developed, such as refractometer, keratometer, corneal mapper, tonometer, fundus camera,
slit lamp, laser scan ophthalmoscope and optical coherence tomography. Among them, the
refractometer that is used for measuring the refractive power of the human eye has the long research
history and various types have been developed. However the efforts to realize more accurate and
precise measurement are still in progress. The wavefront analyzer commercialized in recent years
is an excellent outcome of such efforts. In this paper, a brief account of the developmental history
of various refractometers including the wavefront analyzer is summarized, and the underlying
measurement principle is introduced in the view of optics. Finally, the technical issues that should
be solved for getting better performance are discussed.