OAK

Simultaneous measurement of refractive index and thickness by combining low-coherence interferometry and confocal optics

Metadata Downloads
Abstract
We propose and demonstrate novel methods that enable simultaneous measurements of the phase index, the group index, and the geometrical thickness of an optically transparent object by combining optical low-coherence interferometer and confocal optics. The low-coherence interferometer gives information relating the group index with the thickness, while the confocal optics allows access to the phase index related with the thickness of the sample. To relate these, two novel methods were devised. In the first method, the dispersion-induced broadening of the low-coherence envelop signal was utilized, and in the second method the frequency derivative of the phase index was directly obtained by taking the confocal measurements at several wavelengths. The measurements were made with eight different samples; B270, CaF2, two of BK7, two of fused silica, cover glass, and cigarette cover film. The average measurement errors of the first and the second methods were 0.123 % and 0.061 % in the geometrical thickness, 0.133 % and 0.066 % in the phase index, and 0.106 % and 0.057 % in the group index, respectively. (C) 2008 Optical Society of America.
Author(s)
Kim, SeokhanNa, JihoonKim, Myoung JinLee, Byeong Ha
Issued Date
2008-04
Type
Article
DOI
10.1364/OE.16.005516
URI
https://scholar.gist.ac.kr/handle/local/17402
Publisher
Optical Society of America
Citation
Optics Express, v.16, no.8, pp.5516 - 5526
ISSN
1094-4087
Appears in Collections:
Department of Electrical Engineering and Computer Science > 1. Journal Articles
공개 및 라이선스
  • 공개 구분공개
파일 목록
  • 관련 파일이 존재하지 않습니다.

Items in Repository are protected by copyright, with all rights reserved, unless otherwise indicated.