Purpose: The purpose of this study was to develop a multi-channel NIRS (Near-Infrared
Spectroscopy) system and verify its diagnostic capability for the early detection and improvement of
the accuracy of imaging diagnosis of prostate cancer.
Materials and Methods: We developed a new multi-channel NIRS system using two near-infrared
wavelengths - 785 nm and 830 nm. Each wavelength had 8 channels and shared 2 detectors. A computer
software named Labview (National Instrument, USA) was used to develop GUI (Graphic User
Interface) which can process incoming signals and display them on the same monitor simultaneously.
We applied our NIRS system to the optical phantoms of the prostate gland and prostate cancer to verify
its diagnostic capability.
Results: Relatively lower light signal intensities were noted around three surface channels over
prostate cancer model at both detectors and at both wavelengths, suggesting that light is more absorbed
and/or scattered when passing through these routes reflecting the presence of an inner region
with different optical properties from its surroundings.
Conclusion: We successfully demonstrated the possibility of the NIRS system as a new potential
imaging method for improving the diagnostic accuracy of prostate cancer.
Author(s)
김승섭정대철송현아이송현김재관김법민
Issued Date
2013
Type
Article
URI
https://scholar.gist.ac.kr/handle/local/15724
Publisher
대한의학영상정보학회
Citation
대한의학영상정보학회지, v.19, pp.44 - 50
Appears in Collections:
Department of Biomedical Science and Engineering > 1. Journal Articles
공개 및 라이선스
  • 공개 구분공개
파일 목록
  • 관련 파일이 존재하지 않습니다.

Items in Repository are protected by copyright, with all rights reserved, unless otherwise indicated.