OAK

Development of Intraoperative Near-Infrared Fluorescence Imaging System Using a Dual-CMOS Single Camera

Metadata Downloads
Abstract
We developed a single-camera-based near-infrared (NIR) fluorescence imaging device using indocyanine green (ICG) NIR fluorescence contrast agents for image-induced surgery. In general, a fluorescent imaging system that simultaneously provides color and NIR images uses two cameras, which is disadvantageous because it increases the imaging head of the system. Recently, a single-camera-based NIR optical imaging device with quantum efficiency partially extended to the NIR region was developed to overcome this drawback. The system used RGB_NIR filters for camera sensors to provide color and NIR images simultaneously; however, the sensitivity and resolution of the infrared images are reduced by 1/4, and the exposure time and gain cannot be set individually when acquiring color and NIR images. Thus, to overcome these shortcomings, this study developed a compact fluorescent imaging system that uses a single camera with two complementary metal-oxide semiconductor (CMOS) image sensors. Sensitivity and signal-to-background ratio were measured according to the concentrations of ICG solution, exposure time, and camera gain to evaluate the performance of the imaging system. Consequently, the clinical applicability of the system was confirmed through the toxicity analysis of the light source and in vivo testing.
Author(s)
Choi, JanghoonShin, Jun GeunKwon, Hyuk-SangTak, Yoon-OhPark, Hyeong JuAhn, Jin-ChulEom, Joo BeomSeo, YoungseokPark, Jin WooChoi, YongdooEom, Jonghyun
Issued Date
2022-08
Type
Article
DOI
10.3390/s22155597
URI
https://scholar.gist.ac.kr/handle/local/10679
Publisher
MDPI
Citation
SENSORS, v.22, no.15
ISSN
1424-8220
Appears in Collections:
Department of Biomedical Science and Engineering > 1. Journal Articles
공개 및 라이선스
  • 공개 구분공개
파일 목록

Items in Repository are protected by copyright, with all rights reserved, unless otherwise indicated.